SCOPE BUDDY USER/SERVICE MANUAL
26
Figure 10 Scope Buddy Line Hook-up
Note: Detergent and rinse water cleaning solutions can be drawn from either a clean
container for a single-pass through Scope Buddy, or recirculated from the sink through
Scope Buddy. Scope Buddy can be used in either format depending on the facility’s
endoscope cleaning procedure.
Caution: Verify fluid exiting the open end of the endoscope tubing is not facing a direction
that would allow fluid to be sprayed on personnel or electrical fixtures. The endoscope
should always be fully immersed under water.
Flushing an Endoscope with Detergent
1. Plug the unit in and then turn it on by pressing the power switch located on the side of the unit. The
timer display will now illuminate. See Figure 11.
Figure 11 Base Unit Power Cord Connection
Power Switch
Sco
p
e Budd
y
Unit
Elevator Wire Channel Hook-u
Flow Intake Line