EasyManua.ls Logo

Medivators SCOPE BUDDY - Troubleshooting Guide

Medivators SCOPE BUDDY
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MEDIVATORS REPROCESSING SYSTEMS
33
6. If the dispensed volume is above or equal to 100 ml’s, then SCOPE BUDDY is ready for use. If the
dispensed volume is below 100 ml’s, then contact your local distributor or Medivators Technical
Support. Do not use SCOPE BUDDY if the dispensed volume is below 100 ml’s.
7. Record the test results by marking a Pass or a Fail on the Scope Buddy Daily Quality Assurance Log.
WARNING: Do not use SCOPE BUDDY if it fails the Daily Quality Assurance Test. Doing
so can result in endoscopes which are not flushed with the correct volume of detergent or
rinse water solutions.

Other manuals for Medivators SCOPE BUDDY

Related product manuals