4. RECALL: The RECALL key is provided as a convenience to users and recalls the last
mode and the last power setting used after the system is turned on, and the recall button
is pushed. During use, toggling between modes will maintain the last power setting use.
5. Bipolar: The BIPOLAR controls are color coded blue. The large numerical
display indicates the power setting. The arrow ‘up’ key will increase power
settings in increments of one watt up to 80W. The arrow ‘down’ key will
decrease power settings in increments of one watt.
When the bipolar mode is used the current will flow between the two tips of the
instrument and will desiccate tissue. The bipolar mode can be activated with
either a hand switching device, or with a bipolar footswitch.
The Mega Power offers two bipolar settings, a standard/micro bipolar and a
macro bipolar.
Standard/Micro Bipolar
The Mega Power automatically boots up in the Standard/Micro bipolar setting.
This setting is utilized for the majority of all bipolar cases and delivers a precise
and controlled bipolar effect. The initial voltage is kept low to prevent sparking
and the system maintains a specific power curve to achieve the proper tissue
effect.
Macro Bipolar
Bipolar cases requiring a more rapid onset of power and higher voltage power
output utilize a macro bipolar power curve. To access the Macro bipolar setting,
simultaneously press the Standard and Spray Coag buttons as instructed below.