28
Declaration for Electromagnetic Emissions
The Mega Power is intended for use in the electromagnetic environment specified below. The user of the
Mega Power should assure that the Mega Power is used in such environment.
RF Emissions CISPR 11 Group 2 The Mega Power must emit electromagnetic
energy in order to perform its intended function.
RF Emissions CISPR 11 Class A
Harmonic Emissions IEC 61000-3-2 Class A
Voltage Fluctuations/Flicker Emissions
IEC 61000-3-3
Complies
The Mega Power is suitable for use in all
establishments other than domestic and those
directly connected to the public low-voltage power
supply network that supplies buildings used for
domestic purposes.
Declaration for Electromagnetic Immunity
The Mega Power is intended for use in the electromagnetic environment specified below. The user of the Mega
Power should assure that the Mega Power is used in such environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic environment -
guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
± 8kv air
Full compliance Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%.
Electrical fast
transient/burst
IEC 61000-4-4
± 2kv for power
± 1kv for
input/output lines
Full compliance Mains power quality should be that of a
typical commercial or hospital
environment.
Surge IEC 61000-4-5 ± 1kv differential
± 2kv common
mode
Full compliance Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions and voltage
variations on power supply
input lines.
IEC 61000-4-
<5% Ut (>95% dip in
Ut) for 0.5 cycle
40% Ut (60% dip in
70% Ut (30% dip in
<5% Ut (>95% dip in
Ut for 5 sec
Full compliance Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Mega
Power requires continued operations
during power mains interruptions, it is
recommended that the Mega Power be
powered from an uninterruptible power
supply or a battery.
Power frequency (50/60
Hz) magnetic field.
IEC 61000-4-8
3 A/m Full compliance Power frequency magnetic fields should
be at levels characteristic of a typical
commercial or hospital environment
Disposal
The equipment can be disposed of at the end of its useful life as with normal electronic scrap.
Notice: Follow local governing ordinances and recycling plans regarding disposal or recycling of
device components.