Cautions
The cautions given below pertain to testing, calibration, and servicing of
the generator. The service technician should also be aware of the cautions
given in the user manual for clinical use of the generator
Read all instructions prior to installation and operation.
Use properly grounded AC outlets when connecting the generator power
cord.
Do not place containers of fluids on the generator, or allow fluids to spill on
the generator.
Do not operate the generator without adequate clearance for ventilation.
Space between the bottom of the feet and the floor of the generator should
be free of obstruction. At least two inches of air should separate the sides,
back, and top of the generator from any ventilation-obstructing surface. Do
not place the generator in a small, non-ventilated location.
Connect accessories (e.g. pencil, foot cord, bipolar instruments, return
electrode, etc.) to the proper receptacle using connectors specifically
designed for the intended purpose. Follow the instructions for use provided
by accessory manufacturers.
Apparent low power output or failure of the electrosurgical equipment to
function correctly at normal settings may indicate faulty application of the
dispersive electrode or failure of an electrical lead. Do not increase power
output before checking for obvious defects or misapplication.
Do not coil electrosurgical accessory cords, and do not wrap electrosurgical
accessory cords around metal objects. This may induce current flow to
unintentional areas causing, shocks, burns,
With the fan assembly removed, do not activate the generator for more than
one half hour at full rated power and load. Activating the generator beyond
the recommended Duty Cycle of ten seconds on and thirty seconds off for
long periods of time can cause system overheating.
When repairing the generator, follow the static control guidelines for
semiconductors in order to avoid damaging individual components or board
assemblies.