25
Technical Specifications
General Equipment Specifications
Equipment Classification IEC 60601-1 Class I
Equipment Type IEC 60601-1 Type CF
Degree of protection against ingress of water Meets ANSI/AAMI HF-18,
UL 60601-1 1
st
Ed.,
IEC 60601-2-2
Patient Circuit Isolated from Earth Ground
Cooling Natural Convection,
Modulated Internal Fans
Physical Characteristics
Width 40.5 cm (15.9 inches)
Depth 40.5 cm (15.9 inches)
Height 20 cm (7.9 inches)
Weight 8.65 kg (19 lbs)
Operating Environment
The Mega Power generator is designed and tested to operate within the following
environmental parameters.
Temperature Range: +10°C (+50°F) to +40°C (+104°F)
Humidity Range: 15% to 75% Non-condensing
Atmospheric Pressure: 700 hPA (10.2 psi) to 1060 hPA (15.37 psi)
Warm-up requirements
If the Mega Power is stored outside of the above range, allow the unit to stabilize at
room temperature for a minimum of one hour before use.
Storage Environment
The Mega Power generator is designed and tested for storage within the following
environmental parameters.
Temperature Range: -40°C (-40°F) to +70°C (+158°F)
Humidity Range: 10% to 100 %, Condensing
Atmospheric Pressure: 500 hPA (7.25 psi) to 1060 hPA (15.37 psi)
Audio frequencies
ACE Cut, Pure CUT and Blend 840 Hz ± 10%
Standard Coag and Spray Coag 520 Hz ± 10%
Bipolar 480 Hz ± 10%
Alarms
Power Limit (One beep) 1980 Hz ± 10%
Errors (Two beeps) 1980 Hz ± 10%
CQM (Three Beeps) 2550 Hz ± 10%
The volume of the CUT, COAG and BIPOLAR active signals can be adjusted up and
down. They have a minimum volume of 40 dB at a distance of one meter. The alarm
volume is fixed to exceed 65 dB at a distance of one meter per the requirements of
ANSI/AAMI HF-18.