21. Adjust the variable resistor to increase the resistance until the alarm
sounds and the light turns red. Read the variable resistor, it should be
between 58 and 72 ohms.
Megadyne recommends a yearly calibration check of the Mega Power Electrosurgical
Generator. If your Mega Power is found to be out of calibration, please send your
system directly to Megadyne for calibration. Contact Megadyne customer service to
obtain an RGA number and shipping instructions prior to sending the Mega Power in for
calibration.