www.megger.comDLRO100E, 100X & 100H – 100A
11
100A
50A
10A
MAN
AUTO
Test aborted
1.7 User configured continuous test - Initial
Press test button to start a custom continuous test.
To configure test setting for user configured test, date and time
for saved results - refer to Instrument set-up section 2.6
1.10 Test aborted by instrument
Aborted test - A number of conditions can cause an abort. This
example shows a bad connection.
1.11 Test stopped by user
User stopped test by pressing the test button.
1.8 User configured continuous test - Progress
Active continuous test screen.
100A
50A
10A
MAN
AUTO
End of Test
1.9 End of test screen
Display shows the current through the test piece, measured
voltage and calculated resistance. If the required current is not
achieved the current value will flash.