11. Warning messages
The following warning messages may be displayed during the testing process.
Characters which appear in the aux (small) digit field on the display are shown here in a slightly smaller font size.
11.1 Startup warnings
“UNC” -Instrument is un-calibrated
11.2 Battery
“bAt” -Low battery
11.3 Battery charger
“bAt CHA” -Battery charging
“bAt FUL” -Battery fully charged
11.4 Fuse warning
“FUS” -Fuse blown.
11.5 Invalid rotary switch setting
“ERR - - -“ -General error – invalid combination of rotary switches.
11.6 Continuity test
“VOL 0-L” -Voltage overload during test
11.7 Insulation test
“1000V 1000V” -Flashing warning before 1kV test.
“VOL 0-L” -Voltage overload during test
11.8 RCD Test
“trp” -Supply tripped unexpectedly.
“>50V” -Test aborted due to danger of exceeding touch-voltage limit.
“Err con” -Hardware problem detected during High Current Loop test or RCD test.
11.9 RCD range selection errors
“ERR >1000mA” -Requested current is >1000mA.
“ERR - - -“ + Type A -In Type A mode -Instrument has been set for Type A test, but Type A test is not valid
with this setting.
“ERR - - -“ + Type B -In type B mode
- Instrument has been set for Type B test, but Type B test is not valid with this setting.
“ERR - - -“ + Type S breaker symbol
-Instrument has been set for Type S test, but Type S test is not valid with this setting.
“ERR HI mA” -On VAR range, current is set too high for the selected test.