1 8
T O R K E L 8 2 0 / 8 4 0 / 8 6 0
P r o g r a m m a E l e c t r i c A B
Z P - B S 0 6 E R 1 0 1
7
Testing with a load
profile
Preparations
To configure TORKEL for a profile test, proceed as
follows:
Press <ESC> repeatedly until you see the ”MAIN
MENU”.
Select ”Test method” using the vertical arrow
keys (
or
) and press <ENTER>.
Select ”PROFILE I” for a current profile (or ”PRO-
FILE P” for a power profile) and press <ENTER>.
1]
2]
3]
General
TORKEL can be used to conduct a test that incorpo-
rates a current profile or power profile. A profile can
consist of up to 19 time intervals. The duration and
the magnitude of the load can be specified for each
interval.