AJ0383DE ZP-AJ01E TRAX
55
Inputs
ANALOG
1 2 3 4
Current 4 x 0-10 A AC / DC
Voltage
4 x 250 / 350 V AC / DC
R1 R2 2 x 0-50 V DC Intended for resistance
measurements but can be
used for AC voltage meas-
urement up to 40 V RMS
TRANS Input for analog trans-
ducers and low level analog
signals
TRIG IN Contact or voltage sense
TIMING 3 x 0-10000 s Binary inputs for timing
measurements in timer and
relay testing applications. A
and B inputs dedicated for
Start and Stop.
Calculated / displayed parameters
Arithmetic +, -, *, /
Power P, VA, Q, S
Impedance R (DC), Z, Xp, Xs, Rs, Rp, Ls, Lp, Cs, Cp, phase
Derating at lower mains voltage
TRAX specification is valid at 230-240 V mains voltage. Output
power is decreased at lower mains voltages.
Derating at high ambient temperature
TRAX specification is valid at 23 ±5°C. Max output current times
will be reduced when using TRAX in high ambient temperature.
Derating at lower frequencies
TRAX voltage output specification is at 50 Hz. Maximum voltage
output at lower frequencies is limited by the transformer. Derating
is linear with frequency and max voltage output at 5 Hz is 10% of
rated output.
Measurement accuracy
External
AC / DC volta ge
and current
0.05% of reading + 0.05% FS
Internal DC
current
0.1% of reading + 0.1% FS
Internal AC
current
0.2% of reading + 0.2% FS
Internal AC
voltage
0.2% of reading + 0.2% FS
COM
Ethernet port For running the instrument from an external PC
or connect it to an external network.
Connector for
Wifi antenna
For running the instrument wireless from a PC or
tablet. (Option)
USB 3 USB ports for multipurpose use