22
Make Run Time Changes
To change CH1 frequency: Press CH1, the edit cursor will jump to CH1
frequency value. Then enter the digits of
the new frequency value. For example, to
enter frequency 970 Hz, press CH1, 9, 7,
0, 0 and SET.
To change CH2 frequency: Press CH2, the edit cursor will jump to CH2
frequency value. Then enter the digits of
the new frequency value. For example, to
enter 52.9Hz, press CH2, 5, 2, 9 and
SET.
To change output current intensity +/- : Press + key to increase current
intensity value or press - key to decrease
current intensity value. Or press SET key
until the edit cursor has moved to the
output current position, then use number
keys or arrow keys to change the value.
To change polarity: Press POL key to rotate the output polarity between ±
(alternating), + (polarized positive) and
(polarized negative).
To change waveform: Press WF key to rotate the list of predefined waveforms.
Use UP and DOWN arrow to scroll
forward or backward the waveforms.
To change run time: The run time of each frequency pair is programmed in the
protocol. Press T+ and T keys to increase
and decrease by 1 minute the run time.
Pressing key HLD will hold the timer and
there will still be microcurrent at outputs.
To HOLD current frequency pair: Press
HLD key to hold the timer of current