User Guide Version 2.60 I-button WHO 5th
Revision 21_Sep_2015 20
POSTVASECTOMY
# SPERM/SCAN:
MOTILE > 30
PLEASE RE-RUN AS A NORMAL TEST
POSTVASECTOMY
# SPERM/SCAN: # SPERM/SAMPLE VOL:
MOTILE 3 MOTILE 0.2 M
IMMOTILE 8 IMMOTILE 0.5 M
TOTAL 11 TOTAL 0.7 M
Leave the testing capillary in the
visualization chamber and transfer the
test results to V-Sperm to capture and
attach a video clip of the sample in the
patient’s record.
If the SQA-V reports > 30 motile
spermatozoa, a screen will indicate that a
NORMAL TEST should be run instead of a
POSTVASECTOMY test.
> 30 motile spermatozoa is equivalent to
MSC > 2M/ml.