EasyManua.ls Logo

MES SQA-V Gold - Set-Up: Non Assayed Control; Running CONTROLS on the Automated System

MES SQA-V Gold
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
User Guide Version 2.60 I-button WHO 5th
Revision 21_Sep_2015 22
Please note:
To run 10 replicates:
After each completed
test, remove the
capillary and initiate
the CONTROL test
again using the same
capillary.
Set-Up: Non-Assayed Material to Establish the target value and +/- range
This is also the set-up procedure for sperm concentration proficiency
challenge
Follow the same Steps 1-5 for “Set-up: Assayed Control” above.
Step 6: Enter the following information from the product labeling
LOT# - number identifying the control media lot.
EXP. DATE – control media expiration date (MM=month, YY=year).
Step 7: Enter the TARGET VALUE and +/- Range for Level 1 and Level 2:
Enter 00 for the target value.
Enter 0.0 for the +/- range.
NEGATIVE control target value and +/- range is pre-set to 0.0
Step 8: Save settings: Press APPLY. The set-up takes about two minutes.
Step 9: Establish the target value and +/- range for each level:
Fill a testing capillary and run 10 replicates following the instructions
below “Control Testing.”
Calculate the mean target value. Based on laboratory protocols
determine the +/- range (Example: 2SD).
Follow steps 1-7 of “Set-Up: Assayed Control” to update the target
value and +/- range for the control.
Running
Controls in
the SQA-V
CONTROL Testing
MAIN MENU
Select RUN CONTROLS from the
MAIN MENU of the SQA-V.
TEST NEW PATIENT
RUN CONTOLS
The Control defaults have already
SERVICE
been set-up in V-Sperm.
Select the CONTROL LEVEL: #1, #2
CONTROL LATEX BEADS
or NEGATIVE (LEVEL #3) that is being
SELECT:
tested.
CONTROL LEVEL:
Press ENTER to continue.
LEVEL #1/LEVEL #2/NEGATIVE CONTROL
Controls are run in exactly the same
PRESS ENTER TO CONTINUE
manner as a normal semen sample.
Using control media, follow the same
CONTROL: LATEX BEADS, LEVEL #1
procedure for filling an SQA-V testing
capillary with a NORMAL volume FILL, CLEAN AND WIPE CAPILLARY
sample.
INSERT IN CHAMBER
Testing will begin automatically.
TESTING WILL BEGIN
Control test results will be displayed
AUTOMATICALLY
on the SQA-V screen.
CONTROL TEST RESULTS
LOW, HIGH or NORM. will be displayed
DATE 01/12/06 DD/MM/YY TIME 15:09:08
based on the testing outcomes vs.
LEVEL #1 LOT# 11223344556677889900
target value and +/- range (Disregard
EXP. DATE 04/09 MM/YY
this for non-assayed controls target
TYPE: LATEX BEADS
range set at “0”).
TARGET VALUE: 45.0 +/- 6.3 M/ml
Test results will automatically be saved
CONC. RESULTS: 45.4 M/ml NORM.
and printed.
ACCEPTABLE RANGE: 38.7 – 51.3 M/ml

Table of Contents

Other manuals for MES SQA-V Gold