35
• Clean the cu surface by wiping it with a damp cloth with mild
soap water or with wipes.
• Do not wash the cu or immerse it in water. Do not use petrol,
thinners or similar solvents to clean the cus.
Cleaning – cus:
Do not sterilise cus. Do not use bleach.
Wash and rinse temperatures should not exceed 40° C.
MAINTENANCE
NOTE
Do not wash the cus in a washing machine or iron them.
NOTE
In the event of mechanical stress, the device must be calibrated!
6.3 DISINFECTION
Approved disinfectants:
• Isopropyl alcohol 70%
• Propanol (70-80%)
• Ethyl hexanal
• Aldehyde (2-4%)
• Ethanol (70-80%)
For disinfection, use commercially available disinfectants intended for
the professional healthcare environment. Refer to the Instructions for
Use of the producer.
Not approved disinfectants:
• Organic solvents
• Ammonia-based detergents
• Abrasive cleaning agents
• 100% alcohol, acetone, chlorine, etc.
Disinfection – cu:
• Can be disinfected by 70% medical-grade alcohol.
• After disinfection, the cu cover should be left air dry to
air-dry.