40
ERRORS
ERRORS
8
ERRORS
Abnormally weak pulse
detected. Possibility of
severe PAD or incompress-
ible arteries. Please check
the cuff position and
repeat the measurement.
The pulse was too
weak. Reposition
and repeat the
measurement.
Check the position of the cuff
and repeat the measurement.
Follow Chapter 5.3.1 Cuff
placement.
Error
Description
Solution
Nr of
ERROR
PAD
An anomaly has been
detected. The patient
may have moved during
the measurement
process.
E2
The pulse was too
weak. Reposition
and repeat the
measurement.
Check the position of the cuff
and repeat the measurement.
Follow Chapter 5.3.1 Cuff
placement.
The cuff was insuffi-
ciently inflated.
E3
Inflation
insufficiency.
Check the placement of the cuffs
and repeat the measurement.
An error occurred
during heart rate calcu-
lation.
E4
An error occurred
during heart rate
calculation.
Repeat the measurement. If the
error persists, the measured value
is outside the measurement range
of the device.
An error occurred
during ABI calculation.
E5 An error occurred
during ABI
calculation.
Repeat the measurement. If the
error persists, the measured value
is outside the measurement range
of the device.
Inflation error. Check the
placement of the cuffs
and repeat the measure-
ment.
E6 An error occurred
during the inflation
of the cuff.
Check the placement of the cuffs
and repeat the measurement.
Deflation too fast. The
cuff is not attached or is
not attached correctly.
Place the cuff correctly
and repeat the measure-
ment.
E7
The cuff was
insufficiently
deflated.
Check the placement of the cuffs
and repeat the measurement.
An error occurred during
systolic pressure calcu-
lation.
E8 An error occurred
during systolic
pressure calculation.
Repeat the measurement. If the
error persists, the measured val-
ue is outside the measurement
range of the device.
An error occurred during
diastolic pressure calcu-
lation.
E9
An error occurred
during diastolic
pressure calculation.
Repeat the measurement. If the
error persists, the measured
value is outside the measure-
ment range of the device.
A large pressure fluctu-
ation was detected. The
patient may have moved.
E11 A large pressure
fluctuation was de-
tected. The patient
may have moved.
The result may be incorrect.
Repeat the measurement.
An error occurred
during mean pressure
calculation.
E10 Repeat the measurement. If the
error is repeated, the measured
value is outside the measure-
ment range of the device.
An error
occurred during
mean pressure
calculation.