STANDARD COMPLIANCE
Phenomenon
Electronic discharge
IEC 61000-4-2
Professional healthcare
facility environment
± 8 kV contact
± 2 kV, ± 4 kV, ± 8 kV, ± 15 kV air
3 V/m f)
80 MHz – 2,7 GHz b)
80% AM 1 kHz c)
Enclosure Port
Radiated RF EM
fields a)
IEC 61000-4-3
3 V/m f)
80 MHz – 2,7 GHz b)
80% AM 1 kHz c)
Immunity test levels
Home healthcare
environment
All environments
Phenomenon
Proximity fields
from RF wireless
communications
equipment
IEC 61000-4-3
Rated power
frequency magnetic
fields d) e)
IEC 610004-8
Test frequency (MHz)
Modulation
Test level (V/m)
385 MHz (18 Hz Pulse Modulation) b)
450 MHz (FM+/-5 KHz deviation 1
kHz sine) c)
710 MHz (217 Hz PM) b)
27 V/m
28 V/m
9 V/m
745 MHz (217 Hz PM) b) 9 V/m
780 MHz (217 Hz PM) b) 9 V/m
810 MHz (18 Hz PM) b) 28 V/m
870 MHz (18 Hz PM) b)
28 V/m
1720 MHz (217 Hz PM) b) 28 V/m
1720 MHz (217 Hz PM) b) 28 V/m
1845 MHz (217 Hz PM) b) 28 V/m
1970 MHz (217 Hz PM) b) 28 V/m
2450 MHz (217 Hz PM) b) 28 V/m
5240 MHz (217 Hz PM) b) 9 V/m
5500 MHz (217 Hz PM) b) 9 V/m
5785 MHz (217 Hz PM) b) 9 V/m
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between
the transmitting antenna and the ME EQUIPMENT or ME SYSTEM may be reduced
to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
b) The carrier shall be modulated using a 50 % duty cycle square wave signal.
c) As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used
because while it does not represent actual modulation, it would be worst case.