EasyManua.ls Logo

METER GROUP WP4C - Page 25

METER GROUP WP4C
58 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
WP4C 5 CALIBRATION AND VERIFICATION
Note: The WP4C will automatically shift to precise sampling
mode for the verification/calibration.
The reading should be within ±0.05 MPa of the correct reading
of the KCl standard at that temperature. At 20
C this should
be 2.19 MPa. At 25
C this should be 2.22 MPa.
If your WP4C is reading within 0.05 MPa of the KCl solution,
press Exit and proceed with reading. If it is not, a change in
calibration may have occurred, or the sensor chamber may be
contaminated. For cleaning instructions, see Chapter 10. After
cleaning, repeat these instructions.
If, after cleaning, you consistently get readings that differ from
the correct water potential of the KCl by more than ±0.05
MPa, a change in calibration has probably occurred. Press the
upper right button to move the value up, the lower right but-
ton to move it down. When it is at the correct value for the
verification standard, press the Exit button. The value will be
stored.
Note: This is the only menu where these buttons can change
the calibration, so you will not hurt anything by pressing these
buttons in other menus.
Read the KCl standard again in the normal reading mode. It
should read the proper value.
If after adjusting the calibration and cleaning the chamber you
still are getting incorrect readings when reading the KCl, con-
tact METER at 509-332-5600 for further instructions.
21

Related product manuals