9 CLEANING AND MAINTENANCE WP4C
Figure 10: View of Inside Block
9.2 Cleaning Procedure
Cleaning your WP4C is a multi-step procedure which involves wash-
ing, rinsing, and drying for each specific area as outlined below:
1. Cleaning the Inside Chamber
Note: Be extremely careful not to damage the fan blades in the
chamber. The fan blades are very fragile; if one of them breaks,
your instrument will not work properly. Take extra care when
cleaning this portion.
(a) Remove any debris that may have collected within or
around the sample chamber.
(b) Wrap a new Kimwipe around the end of the thin plas-
tic rod (spatula) and moisten it with isopropyl alcohol or
cleaning solution.
Note: Do not dip a used Kimwipe into your container