sample size of 1 g when running the MCC standard test. An error value less than 0.12 would
be acceptable. If the error value is greater or equal to 0.12, contact Customer Support for
additional assistance.
If you disconnect the Moisture Analysis Toolkit software at any time while running the
standard test, it does not automatically generate a comparison. Instead, save the completed
MCC isotherm curve as a .vsa file and then manually compare it to the preloaded standard
MCC curve. To do this, select Standard Comparison from the action menu in Moisture
Analysis Toolkit software to open a file dialog box. Navigate to and open the .vsa file for the
MCC isotherm curve. The software outputs this curve at the conclusion of the standard
test and you can compare these curves with any previously saved MCC standard isotherm
curves. Doing so initiates a comparison of this curve to the preloaded standard MCC curve
and the system prompts you to save a .pdf of the comparison report. Customers can use the
standard comparison feature anytime to make a comparison between previously saved MCC
standard isotherm curves and the preloaded MCC standard curve.