31
WP4C
APPENDIX B. TEMPERATURE CORRECTION OF METER
VERIFICATION STANDARDS
Table 6 provides the temperature correction for multiple verification standards available
from METER.
Table 6 Temperature correction of verification standards
Temperature
(°C)
Water
0.50 mol/kg
KCl
2.33 mol/kg
NaCl
6.00 mol/kg
NaCl
8.57 mol/kg
LiCl
13.41 mol/kg
LiCl
17.18 mol/kg
LiCl
15 0 -2.14 -10.66 -36.32 -94.32 -190.89 -261.45
20 0 -2.18 -10.99 -37.13 -94.86 -190.28 -261.24
25 0 -2.22 -11.47 -37.76 -95.37 -190.75 -261.03
30 0 -2.26 -11.67 -38.39 -95.86 -191.18 -260.83
35 0 -2.29 -11.86 -39.03 -96.32 -191.02 -260.61
40 0 -2.33 -11.89 -39.66 -96.74 -191.38 -260.39
50 0 -2.41 -16.71 -44.90 -98.38 -192.53 -262.51