MI 3205 TeraOhm XA 5kV Technical specifications
56
Current
Measuring range (A) Resolution (A) Accuracy
1.00 5.00 m 10
(5 % of reading + 3 digits)
100 999 1
10.0 99.9
100 n
1.00 9.99
10 n
100 999 n
1 n
10.0 99.9 n
100 p
0.00 9.99 n
10 p
(10 % of reading + 0.15 nA)
Table 8.2: Current measurement ranges and accuracy
Voltage
Measuring range (V) Resolution (V) Accuracy
0 999
1
(5 % of reading + 3 digits)
1.00 5.99 k
10
Table 8.3: Voltage measurement ranges and accuracy
Capacitance
Measuring range (F) Resolution (F) Accuracy
20 999 n
1 n
(5 % of reading + 3 digits) 1.00 9.99
10 n
10.0 50.0
100 n
Table 8.4: Capacitance measurement ranges and accuracy
Nominal voltage range ................................... 500 V 5 kV
Dielectric absorption ratio DAR
Display range DAR Resolution Accuracy
0.01 9.99
0.01
(5 % of reading + 3 digits)
10.0 100.0
0.1
Table 8.5: DAR display ranges and accuracy
Polarization index PI
Display range PI Resolution Accuracy
0.01 9.99
0.01
(5 % of reading + 2 digits)
10.0 100.0
0.1
Table 8.6: PI display ranges and accuracy