3.2. Remote control commands
794 Basic Titrino
124
.Right: Relative specification of the width of the output medium (e.g. pa-
per width) for the length of the measured value axis. 1 means the
measured value axis is plotted over the entire width of the paper
(largest possible width). In extreme cases, the writing of the right
tick may lie outside.
.Feed: Length of the volume axis referred to the burette cylinder volume,
V(B) per cm (0.1 means, e.g. 1 mL/cm with a 10 mL Exchange
Unit). Depending on the printer, the measure in cm may not al-
ways be correct.
3.2.2.99. Setup.PowerOn $G
Simulation of 'power on'. The device has the same status as after power on:
The cylinder is filled, error messages deleted and the current sample number
set to 0. The method last used is ready for operation.
3.2.2.100. Setup.Initialise $G
Setup.Initialise.Select ActMeth, Silo, Calib, Config,
Assembly, Setup, All
Setting of default values for the following areas:
ActMeth: Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
Silo: The silo memory is deleted. Same function as delete entire silo.
Calib: pH calibration data for all measuring inputs.
Config: All values under &Config.
Assembly: All values under &Assembly.
Setup: All values under &Setup.
All: Values of the entire tree (except silo and method memory).
The action must be triggered with &Setup.Initalise $G.
3.2.2.101. Setup.RamInit $G
Initializes instrument, see page 147. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted. The user memory contains the default user methods from Metrohm.
3.2.2.102. Setup.InstrNo $G
Setup.InstrNo.Value serial number, 8 ASCII characters
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
3.2.2.103. Diagnose.Report $G
Output of the report containing the adjustment parameters. The Titrino has to
be in its inactive basic state.