EasyManuals Logo

Metrohm 794 Basic Titrino User Manual

Metrohm 794 Basic Titrino
198 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #172 background imageLoading...
Page #172 background image
6.3. User methods
794 Basic Titrino
168
6.3.4 "Chloride"
Electrode:
6.0430.100 Ag-Titrode at measuring input 1.
Titrant:
c(AgNO
3
) = 0.01 mol/L for low Cl
-
contents,
e.g. in tap water
c(AgNO
3
) = 0.1 mol/L for higher Cl
-
contents,
e.g. in food
Sample:
Dissolve sample and add HNO
3
.
Remarks:
Select the appropriate formula. The other may
be deleted.
Reference:
METROHM Application Bulletin Nr. 130
Content of Cl
-
in ppm
Content of table salt in %
Sample size in g
Concentration of titrant
Molar mass of Cl
Factor for ppm
Molar mass of NaCl
Concentration of titrant
Factor for %
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 09:40 0
DET U Chloride
parameters
>titration parameters
meas.pt.density 4
min.incr. 10.0 µl
titr.rate max. ml/min
signal drift 50 mV/min
equilibr.time 26 s
start V: OFF
pause 0 s
meas.input: 1
temperature 25.0 °C
>stop conditions
stop V: abs.
stop V 99.99 ml
stop U OFF mV
stop EP 1
filling rate max. ml/min
>statistics
status: OFF
>evaluation
EPC 5
EP recognition: all
fix EP1 at U OFF mV
pK/HNP: OFF
>preselections
req.ident: OFF
req.smpl size: all
activate pulse: OFF
------------
'fm
794 Titrino 01102 794.0010
date 2002-01-03 time 09:40 0
DET U Chloride
>calculations
Chloride=EP1*C01*C02*C03/C00;2;ppm
NaCl=EP1*C04*C05*C06/C00;2;%
C00= 1.0
C01= 0.01
C02= 35.45
C03= 1000
C04= 58.44
C05= 0.1
C06= 0.1
------------

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 794 Basic Titrino and is the answer not in the manual?

Metrohm 794 Basic Titrino Specifications

General IconGeneral
BrandMetrohm
Model794 Basic Titrino
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals