2.17 Storing determination results and silo calculations
794 Basic Titrino
73
2.17 Storing determination results and silo
calculations
2.17.1 Storing determination results
If the sample-specific data of the silo memory should be kept after the determination and
supplemented by results, the following entries are necessary:
1. In the method under <DEF>
Assignment of the determination results to C24 and/or C25:
2. In the silo memory, <SMPL DATA> (when the silo memory is switched on):
"save lines: on"
Assignment of determination results
The determination results are assigned in key <DEF>.
The display texts of the Titrino are shown to the left.
The values are the default values.
>silo calculations
C24=
C25=
Silo calculations
Assignment to C24 (RSX, EPX, CXX)
Calculated results (RSX), endpoints (EPX) or variables
CXX can be stored as C24.
Same procedure for C25.
Important:
Ensure that there is still sufficient space for storing the results C24 and C25. (In the report
<PRINT><USER METH><ENTER> the number of free bytes is shown.) Result name,
value and unit are stored. The memory requirements can be estimated as follows:
Result with text (8 characters) and unit (5 characters): 32 bytes
Measured value C40, value without unit: 22 bytes
DEF
(
2
def
>silo calculations

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 794 Basic Titrino and is the answer not in the manual?

Metrohm 794 Basic Titrino Specifications

General IconGeneral
BrandMetrohm
Model794 Basic Titrino
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals