2.12. TIP, Titration procedure
794 Basic Titrino
62
2.12 TIP, Titration procedure
TIP (Titration Procedure) is used to link several commands in a sequence.
TIP is selected with <MODE> and <ENTER>. TIP is an "empty shell" in which the
sequence of the determination must be defined.
Definition of the sequence
With key <DEF> the TIP can be defined.
sequence:
Sequence of TIP.
The display texts of the Titrino are shown to the left.
The values are the default values.
>sequence
1.step: OFF
<ENTER>
1.method: Chloride
<ENTER>
etc.
2 x <QUIT>
Sequence
Select a step with keys <SELECT>:
method: Method from the user memory.
pause: Waiting time
L4, L6 output: Set an output.
info: Hold sequence and write a message
into the display.
stirrer ON/OFF
Confirm the step with <ENTER> and enter the
parameter for the selected step, see below.
The request for the second step follows etc. Up to 30
steps can be selected.
On completion of the sequence definition, exit the
inquiry with <QUIT>.