6.3. User methods
794 Basic Titrino
170
6.3.6 "Diazo"
Diazotation of sulfonamides and primary
amines.
Electrode:
6.0341.100 Pt-Titrode, at measuring input 1
Titrant:
c(NaNO
2
) = 0.1 mol/L.
Solvent:
w(HBr) = 0.3 (30%)
Sample:
Dilute 0.2 – 0.5 mmol of the substance to be
determined in 30 mL solvent and add 20 mL
dist. water. Titrate immediately.
– Sample size in g
– Molar mass of the substance to be determined
– Normality of the titrant (0.1*titer)
– Factor for %
'pa
794 Titrino 01102 794.0010
date 2002-01-03 time 09:43 0
MET U Diazo
parameters
>titration parameters
V step 0.10 ml
titr.rate max. ml/min
signal drift OFF mV/min
equilibr.time 20 s
start V: abs.
start V 0.5 ml
dos.rate max. ml/min
pause 80 s
meas.input: 1
temperature 25.0 °C
>stop conditions
stop V: abs.
stop V 6.00 ml
stop U OFF mV
stop EP 9
filling rate max. ml/min
>statistics
status: OFF
>evaluation
EPC 30 mV
EP recognition: greatest
fix EP1 at U OFF mV
pK/HNP: OFF
>preselections
req.ident: OFF
req.smpl size: value
activate pulse: OFF
------------
'fm
794 Titrino 01102 794.0010
date 2002-01-03 time 09:43 0
MET U Diazo
>calculations
Content=EP1*C01*C02*C03/C00;2;%
C00= 1.0
C01= 1
C02= 1
C03= 0.1
------------

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Metrohm 794 Basic Titrino and is the answer not in the manual?

Metrohm 794 Basic Titrino Specifications

General IconGeneral
BrandMetrohm
Model794 Basic Titrino
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals