4.1 Error and special messages
794 Basic Titrino
133
4 Error messages, Troubleshooting
Data transfer inoperative See measures on page 131.
4.1 Error and special messages
XXX bytes missing XXX bytes missing. For the storage of a method or a silo line
XXX bytes are missing or there is insufficient RAM for a TIP
sequence.
Remedy: <QUIT>. Delete methods no longer needed or use
fewer silo lines.
check electrode
With polarized electrodes. There is a break or short circuit.
Possible causes and rectification of the fault:
- the electrode is not plugged in plug it in
- the electrode is not immersed in the solution immerse it
- the electrode is defective use new electrode.
- the electrode cable is defective use new cable.
The electrode test can be switched off under the <PARAM>
key.
Exit: Rectify fault or <STOP>.
check exchange unit The Exchange Unit is not mounted (properly).
Exit: Mount Exchange Unit (properly) so that the coupling
engages or <STOP>.
check T-sensor No temperature sensor is attached (with MEAS T or if the
temperature monitoring is active).
Exit: Connect Pt100 or Pt1000 or <STOP>.
data set reevaluation Message if Titrino is reevaluating downloaded measuring point
lists.
division by zero The result could not be calculated as a divisor in the formula
was equal to zero.
Exit: Enter appropriate value.
EP overflow In a DET or MET titration 9 EP's or more were found. The first 9
EP's will be listed.
Remedy: Recalculation of data with higher EP criterion.
manual stop The determination has been manually stopped.
meas.pt list overflow Maximum 500 measured points can be stored.
Exit: Use start criteria or select larger time interval.
missing EP An EP needed for calculation in a formula is missing.
missing fix EP A fix EP needed for calculation in a formula is missing.
no.EP not corresponding In DET or MET titrations, the number of EP's actually found
does not match the set windows: Exactly 1 EP per window has
not been found.
no EP set In SET, no EP has been set.
Exit: <STOP> and set EP.
no meas.quantity In TIP a manual measurement (<MEAS/HOLD>) has been
performed without defining a measuring quantity.
Exit: <MEAS/HOLD> and define measured quantity.