3.2 Remote control commands
794 Basic Titrino
89
*Parameter Tree part "Parameters for SET"
.SET1 Control parameters for EP1
.EP Endpoint 1 depends on meas.quant. 3.2.2.27.
.UnitEp Unit of endpoint read only ditto
.Dyn Dynamics depends on meas.quant. 3.2.2.28.
.UnitDyn Unit of dynamics read only ditto
.MaxRate Maximum dosing rate 0.01...10...150, max. ditto
.MinRate Minimum dosing rate 0.01...25.0...9999 ditto
.Stop Titration stop
.Type Type of stop criterion drift, time 3.2.2.29.
.Drift Stop drift 1...20...999 ditto
.Time Switch-off delay time 0...10...999, inf ditto
.StopT Stop time 0...999 999, OFF ditto
.SET2 Control parameters for EP2, as for EP1
.TitrPara Titration parameters
.Direction Titration direction +, -, auto 3.2.2.30.
.Start V Start volume
.Type Type of start volume abs., rel., OFF 3.2.2.15.
.V Volume for absolute start volume 0...999.99 ditto
.Factor Factor for relative start volume 0...±999 999 ditto
.Rate Dispensing rate for start volume 0.01...150.0, max. ditto
.Pause Waiting time after start volume 0...999 999 3.2.2.16.
.MeasInput Measuring input 1, 2, diff. 3.2.2.17.
.Ipol Polarization current 0...1...±127 ditto
.Upol Polarization voltage 0...400...±1270 ditto
.PolElectrTest Test for polarized electrodes ON, OFF ditto
.Temp Titration temperature -170.0...25.0...500.0 3.2.2.18.
.StopCond Stop conditions
.VStop Stop volume
.Type Type of stop volume abs., rel., OFF 3.2.2.19.
.V Volume for absolute stop volume 0...99.99...9999.99 ditto
.Factor Factor for relative stop volume 0...±999 999 ditto
.FillRate Filling rate 0.01...150.0, max. 3.2.2.22.
.Statistics Statistics
.Status Status of statistics calculation ON, OFF 3.2.2.23.
.MeanN No. of individual determinations 2...20 ditto
.ResTab Result table
.Select original,delete n,delete all ditto
.DelN Deletion of individual results 1...20 ditto
.Presel Preselections
.Cond Conditioning ON, OFF 3.2.2.31.
.DriftDisp Display of drift during cond. ON, OFF ditto
.IReq Request of Id's after start id1, id1&2, all, OFF 3.2.2.25.
.SReq Request of smpl size after start value, unit, all, OFF ditto
.ActPulse Output of a pulse first, all, cond., OFF 3.2.2.26.