4.7 Menu dialogs
■■■■■■■■■■■■■■■■■■■■■■
52
■■■■■■■■
912/913/914 Meter
Slope upper
limit
Editing dialog for entering the upper limit value.
■ Default value: 103.0 / input range: 1.0 - 999.9%
Limit value
pH(0)
Selection dialog for selecting whether the limit value is to be applied.
■ on
■ off: is the default value
pH(0) lower limit Editing dialog for entering the lower limit value.
■ Default value: 6.40 / input range: 0.00 - 99.99
pH(0) upper
limit
Editing dialog for entering the upper limit value.
■ Default value: 8.00 / input range: 0.00 - 99.99
Limit value c Selection dialog for selecting whether the limit value is to be applied.
■ on
■ off: is the default value
c lower limit Editing dialog for entering the lower limit value.
■ Default value: 0.400 /cm / input range: 0.001 - 500 /cm
c upper limit Editing dialog for entering the upper limit value.
■ Default value: 0.550 /cm / input range: 0.001 - 500 /cm
Temperature off-
set
Editing dialog for entering/modifying the Temperature offset.
■ Default value: 0.0 °C
■ Input range: –5.0 - +5.0 °C
LED intensity
(%)
Selection dialog for selecting the LED intensity (%).
■ 10 %
■ 20 %: is the default value
■ 30 %
Cap type Display field for the Cap type.
■ e.g. MA7-530-200
Limit value dphi
100 %
Selection dialog for selecting whether the limit value is to be applied.
■ on: is the default value
■ off
lower limit Editing dialog for the lower limit.
■ Default value: 15 ° / input range: 15.0 - 30.0 °