4 Operation
912/913/914 Meter
■■■■■■■■
51
Temp. calibra-
tion
Display field indicating the measurement method for temperature measurement
of the last calibration.
Cal. dphi 100 % Editing dialog for entering/modifying the Cal. dphi 100 %.
■ Input range: 15.000 - 30.000
Cal. dphi 0 % Editing dialog for entering/modifying the Cal. dphi 0 %.
■ Input range: 45.000 - 60.000
Cal. temp. 100
%
Editing dialog for entering/modifying the Cal. temp. 100 %.
■ Input range: 0.000 - 99.999 °C
Cal. temp. 0 % Editing dialog for entering/modifying the Cal. temp. 0 %.
■ Input range: 0.000 - 99.999 °C
Calibr. press. Editing dialog for entering/modifying the Cal. dphi 100 %.
■ Input range: 300.000 - 1200.000 mbar
Calibration date
Display field for the last Calibration date.
Calibration time
Display field for the last Calibration time.
Calibration inter-
val
Editing dialog for entering time in days for the Calibration interval.
■ Default value: off / input range: 1 - 999 d
■ off disables the Calibration interval.
Temp. sensor
Selection dialog to indicate the temperature sensor type for the respective sen-
sor.
■ Pt1000: is the default value
■ NTC
– R(25°C)
Editing dialog
Default value: 30000 Ω / input range: 10000 - 100000 Ω
– B value
Editing dialog
Default value: 4100 Κ / input range: 1000 - 9999 Κ
Slope limit value Selection dialog for selecting whether the limit value is to be applied.
■ on: is the default value
■ off
Slope lower limit Editing dialog for entering the lower limit value.
■ Default value: 95.0 / input range: 1.0 - 999.9%