XTP601 User’s Manual
Michell Instruments 11
OPERATION
2.5.2 Chart Page
t
t-4
25
%
0
Figure 11
Chart Page
NOTE: This data is not available via the Modbus
This indicative chart is continuously running at the set chart interval (2
to 60 seconds).
Chart duration in seconds = (chart interval * 60).
It is auto ranging with an auto range resolution of 1% O
2
.
It is reset if the chart interval is changed, the Display is changed between
% O
2
and % O
2
HSR or the instrument power is cycled.
The chart will record either % O
2
(Non HSR) or % O
2
HSR if this is set to
ON in the setup pages.
Chart buffer size is 60 values and therefore the chart duration changes
according to interval chosen.
Chart data is only stored in volatile memory and therefore is not saved.
Chart interval is saved and available in a Modbus register.
Chart data is not available via serial comms as the Application Software is
able to perform more sophisticated charting functions.