APPENDIX C
Appendix C Hazardous Area Certifi cation
The XTP601-EX Oxygen Analyzer is certified compliant to the ATEX Directive (2014/34/
EU) and IECEx for use within Zone 1 & 2 Hazardous Areas and has been assessed so by
SIRA Certification (Notified Body 0518).
The XTP601-EX Oxygen Analyzer is certified compliant to the North American Standards
(USA and Canada) for use within Class I, Division 1, Groups B, C and D Hazardous
Locations and has been assessed so by cCSAus.
C.1 Product Standards
This product conforms to the Standards:
EN60079-0:2009 IEC60079-0:2007
EN60079-1:2007 IEC60079-1:2007
EN60079-31:2009 IEC 60079-31:2008
CSA C22.2 No 0-10 UL 916
CSA C22.2 No 30-M1986 (R 2007) UL 1203
CSA C22.2 No 142-M1987 (R 2009)
C.2 Product Certifi cation
This product is attributed with the product certification codes:
ATEX & IECEx
II 2 GD Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
North American
Class I, Division 1, Groups B, C, D
C.3 Global Certifi cates/Approvals
ATEX Sira12ATEX1042X
IECEx SIR12.0013X
cCSAus 2541173
TC TR Ex RU C-GB. 05.B.00152
INMETRO NCC 14.03166 X
These certificates can be viewed or downloaded from our website at:
http://www.michell.com
C.4 Special Conditions
1. Cable entry holes shall be fitted with either an appropriately certified
cable gland or appropriate certified blanking element. These shall
provide and maintain a minimum enclosure ingress protection of IP66.
2. The maximum pressure associated with the process medium in the
internal pipes shall be limited to 4 bar.
3. The maximum temperature associated with the process medium shall
be limited to 55°C.
4. The flameproof joints of the flame arrestors & breather are not intended
to be repaired.