XTP601 User’s Manual
26 97313 Issue 5, May 2018
INSTALLATION
4 INSTALLATION
Before installing the analyzer, read through this manual carefully and take note of all
warnings.
Do not open the casing while powered-up in a hazardous
area without a 'hot works' permit.
Special care must be taken when the analyzer is
energized.
The XTP601-EX versions must only be installed by
suitably qualifi ed personnel and in accordance with the
instructions provided and the terms of the applicable
product certifi cates.
Maintenance and servicing of the product must only be
carried out by suitably trained personnel or returned to an
approved Michell Instruments’ Service Center.
Before the cover is refi tted, the fl amepath/threaded joint
between the cover and body must be thoroughly wiped
clean of dirt, grit or other foreign substances and then a
thin coating of an approved form of non-setting grease
applied to joint/threads. Ensure gasket is free from
damage.
Threaded covers must be screwed on to a minimum of 7
full threads and then locked in position with the hex key
provided.
The terms of the Hazardous Area Certifi cation do not
permit oxygen measurement in ambient areas with
enriched oxygen.