XTP601 User’s Manual
32 97313 Issue 5, May 2018
INSTALLATION
4.5.3 Signal Output
There are two 4-20mA linear signal output channels. Both are for oxygen concentration.
One is fixed on the calibrated range of the unit and the second can be configured
in the menu. NOTE: When the instrument is warming up (cell temperature
not stabilized) these outputs are factory set to 3.2 mA to indicate that the
instrument is not ready.
The maximum mA output is approximately 20.5mA
The minimum mA output is approximately 3.8mA
The user can select the falut conditon to drive the mA output Low (3.2mA)
or Hi (21.4mA).
4.5.4 Serial Output
The analyzer has Modbus RTU communications over RS485 protocol; please see
Application Software CD for more details.
Type: Modbus RTU over RS485
RS485: 2 wire (plus ground), half duplex
Baud Rate: 9600
• Parity: None
Data bits: 8
Stop bits: 1
4.5.5 Analog (4-20 mA) Outputs and Communications (PL5 - Green)
Warning: Do not attempt to loop-power this instrument via the 4-20 mA
output as this will irreversibly damage the main pcb.
PIN 7 PIN 6 PIN 5 PIN 4 PIN 3 PIN 2 PIN 1
RS485
GND
RS485
B
RS485
A
Ch2
O/P
-
Ch2
O/P
+
Ch1
O/P
-
Ch1
O/P
+
NOTE 1: Channel 1 is fixed range output over instrument range and Channel
2 is adjustable between 0-100%.
NOTE 2: For long cable lengths, a termination resistor of 120Ω is provided to
aid in impedance matching of the communications cable. This is selectable
via JMP3. A terminating resistor is simply a resistor placed at the extreme
end or ends of a cable on an RS485 network to reduce effects of impedance
mismatch. Impedance mismatch can cause reflections of data as it passes
down the cable and the reflections could be large enough to cause data errors.
NOTE 3: The maximum loop load resistance for mA outputs is 550Ω
NOTE 4: the mA outputs are powered with a 24V excitation voltage.