XTP601 User’s Manual
Michell Instruments 25
CALIBRATION
10. The Adjusted reading will now be the same as that displayed on the
Main Page and be equal to the upper calibration gas.
11. The calibration process is complete. Return to sampling the process
gas.
NOTE: The analyzers leave the factory with the field calibration function
turned off. When first calibrating the analyzer go to the settings page and
turn Field Calibration on. This can be done either just before or just after the
field calibration procedure.
3.3 Field Calibration Reset
The Field Calibration can simply be turned ON or OFF in the settings page. But if the
user would like to start again, the Field Cal (including saved data) can be deleted in this
menu.
This feature is accessed by selecting the Reset Page (see below).
DELETE?
FIELD CAL
O2 MIN/MAX
RESET?
DELETE?ALARM LOGS
Figure 27
Field Calibration Reset Page
Select Field Calibration and highlight DELETE?, then press the Up () button 3 times
and then press
ENTER to accept the change.