XTP601 User’s Manual
Michell Instruments 13
OPERATION
2.5.5 Alarms Log Pages
ALARM DATE TIME P1
AL1 02/01 12:50:40
AL1 02/01 11:10:32
AL1 02/01 11:00:29
AL2 02/01 10:20:00
Figure 14
Alarms Log Page
A maximum of 40 High/Low alarms, along with date and time of occurrence, are
recorded in a ring buffer in NV memory. The most recent alarm will overwrite the oldest
alarm when more than 40 alarms are recorded. The data is displayed over a maximum
of 10 pages (with 4 alarms on each page). The latest alarm record is displayed in line
1 of page 1. This data is not available via serial communications or in the blind unit.
The data is reset manually via the Reset Page in the User Menu. The data is saved and
restored when the instrument is restarted. P1 = Page 1.
2.6 Info Page
Firmware Ver
Hours Used
Last Cal Date
Cal Press. mB
Atm Press. mB
ModBus Rx Code
1.0
125
04:08:11
1000.0
1000.0
---
Figure 15
Info Page
From the Main Page it is possible to get to the Info Page by pressing the ESC button.
The information available is displayed below:
Parameter Description
Firmware Ver
Indicates the firmware version installed in the instrument
Hours used
Indicates the number of hours that the instrument has been
powered up
Last Cal Date
The date of the last field or Michell Calibration
Character indicates
F for Field and M for Michell
Cal Press, mB
The atmospheric pressure recorded during the last calibration (used
as null reference point for pressure compensation)
Atm Press, mB
Current atmospheric pressure being read within the analyzer
ModBus Rx Code
The received Modbus function code is flashed here as soon as
a function code is received – this is useful to check the Modbus
communications to ensure that good data is coming through.
If no code is received then ‘
---’ is displayed