XTP601 User’s Manual
Michell Instruments 29
INSTALLATION
4.3 Set-Up
The XTP601 is designed to be panel or wall mounted. There are 2 bolt
holes and 2 lugs (1 per corner) see Figure 11. Dimensional drawings can
be found in Appendix B. Mount the analyzer before attempting to remove
the lid.
WARNING: This unit is 24 V DC powered only!
Do not attempt to loop-power this instrument via the 4-20
mA output as this will irreversibly damage the main pcb.
The EXd thread is greased and therefore it may be advisable to wear latex
gloves.
Ensure that the grub screw is loose, using the hex key (supplied), to
prevent scoring on the casing.
Connect the earthing strap to the earth point on the right hand side ofthe
case.
Remove the lid, using the lid keys (supplied). The lid will require a firm
grip to loosen.
M5 Locking Grub Screw
g
g
g
Figure 29
XTP601 Lid Removal
If fitted, remove the display / status pcb via the two ¼ turn quick-connect
fittings.
Disconnect the ribbon / interconnecting cable from the loose pcb.
WARNING: Prior to connecting power, ensure the unit is
correctly earthed via the earth point on the right hand side
of the case.
Connect to the power and outputs (see Section 4.5).
After making all necessary connections, reconnect the pcb via the ribbon/
interconnecting cable & quick-connect fittings and replace the lid, ensuring
at least 7 full turns on the thread to complete the gas path.
Use the hex key (supplied) to tighten the grub screw. NOTE: This is
necessary in order to adhere to the hazardous area certification
requirements.
For operating instructions refer to Section 2.