EasyManua.ls Logo

Michell Instruments XTP601 - External Compensation Page; Figure 22 External Compensation Page

Michell Instruments XTP601
64 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
XTP601 User’s Manual
18 97313 Issue 5, May 2018
OPERATION
2.7.6 External Compensation Page
COMP 20%
0.50-2.00
0.50-2.00
0.50-2.00
0.50-2.00
0.50-2.00
COMP 40%
COMP 60%
COMP 80%
COMP 100%
Figure 22
External Compensation Page
A 4-20 mA sensor may be used to compensate the % O
2
reading for the effects of
process variables such as line pressure, flow, etc. The table of compensation factors
may be edited for 5 points along the compensation sensor range. The values would be
determined by applying the process variable at each point and noting the effect on the
% O
2
.
For example: a compensation is needed for line pressure. A 4-20 mA line pressure
sensor would be ranged over the compensation range. While the instrument reads a
fixed % O
2
value, a table is created (see example below) while varying the pressure at
20% of range intervals:
Pressure
% of
Pressure
span
O
2
reading
Effect =
(affected value
/ non affected
value)
Compensation
factor = 1 /
effect
0 0% 20.91 20.91/20.91=1.00 1.00
1 20% 21.65 21.65/20.91=1.04 0.96
2 40% 23.56 1.13 0.88
3 60% 25.99 1.24 0.81
4 80% 29.66 1.42 0.70
5 100% 38.85 1.86 0.54
The compensation factor values are then entered into the External Compensation table
(excluding the 0% point as this will always be assumed to be 1 = no effect).
Below 0% (< 4 mA), the compensation factor is fixed to 1. Above 100% the compensation
factor is extrapolated beyond the last factor.

Table of Contents

Related product manuals