XTP601 User’s Manual
20 97313 Issue 5, May 2018
OPERATION
2.7.8 Field Cal Page
CAL TYPE
1/2 POINT
0.00-100.00
0.00-100.00
0.00-100.00
0.00-100.00
**.**
REF GAS 1
ACTUAL 1
REF GAS 2
ACTUAL 2
Adjusted ~
Figure 24
Field Cal Page
Setting Description/Operation
CAL TYPE
1 POINT or 2 POINT
REF GAS 1
Cal reference gas for 1 point cal, lower cal reference gas for 2 point cal
ACTUAL 1
Actual measured O
2
for REF GAS 1
See Section 3.1
REF GAS 2
Upper cal reference gas for 2 point cal
Disabled if 1 point cal selected
ACTUAL 2
Actual measured O
2
for REF GAS 2
See Section 3.2
Disabled if 1 point cal selected
ADJUSTED ~
O
2
reading before and after change
~ symbol will be visible until reading is stable
See Section 3 for field calibration procedure.