XTP601 User’s Manual
Michell Instruments 37
APPENDIX A
Operating Conditions
Ambient Temperature
5 to +55°C (+41 to +131°F)
Atmospheric Pressure
750mbar to 1250mbar
Mechanical Specifi cation
Warm Up Time < 25 minutes
Stabilization Time 5 minutes
Dimensions 234 x 234 x 172mm (9.2 x 9.2 x 6.7”) (w x d x h)
Weight 9.7kg (21.4lbs)
Wetted Materials
316 & 430F stainless steel, boroscillicate glass, platinum, 3M
2216 plus O-ring
O-Ring Materials Viton, Silicone or Ekraz
Gas Connection
EX1, EX3 & GP2: 1/8" NPT (female)
GP1 and GP3: 1/4" NPT (female)
Ingress Protection IP66, NEMA 4X
Hazardous Area Certifi cation
Certification Codes
ATEX II 2GD Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
IECEx Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
INMETRO Ex d IIB+H2 T6 Gb
Ex tb IIIC T85°C Db IP66
TC TR Ex 1Ex d IIB+H2 T6 Gb
cCSAus Class I, Division 1, Groups B C D, T6
Temperature Ranges as per
O-Ring type
Silicone: Ta = -40 to +55°C
Viton: Ta = -15 to +55°C
Ekraz: Ta = -10 to +55°C
The XTP601 process oxygen analyzer meets or exceeds all relevant clauses in BS EN
50104 : 2010 “Electrical apparatus for the detection and measurement of oxygen”.