XTP601 User’s Manual
Michell Instruments 41
APPENDIX B
Addr Function Access Ranges/Resolution Type
40 Cal 02 Ref3 R 0.00-100.00 G
41 Cal 02 Ref4 R 0.00-100.00 G
42 Cal 02 Ref5 R 0.00-100.00 G
43 Cal 02 ADC1 R 0-8191 A
44 Cal 02 ADC2 R 0-8191 A
45 Cal 02 ADC3 R 0-8191 A
46 Cal 02 ADC4 R 0-8191 A
47 Cal 02 ADC5 R 0-8191 A
48 mAINPUT1 4mA Cal point R 0-8191 A
49 mAINPUT1 20mA Cal point R 0-8191 A
50 mAINPUT2 4mA Cal point R 0-8191 A
51 mAINPUT2 20mA Cal point R 0-8191 A
52 mAOUTPUT1 4mA Cal point R 0-8191 A
53 mAOUTPUT1 20mA Cal point R 0-8191 A
54 mAOUTPUT2 4mA Cal point R 0-8191 A
55 mAOUTPUT2 20mA Cal point R 0-8191 A
56 CH2 output zero R/W O2 range min to 0-100%, 0.01 G
57 CH2 output span R/W
CH2 output zero to instrum range
max, 0.01
G
58 Last Cal Date DATE/MONTH R 1-31/1-12 J
59
Last Cal Date: Field or Factory (bit 15) /
YEAR (bits0-3)
R 0=Factory, 1=Field / 0-99 J
60
PCB Tempr Offset (for MSP430 Int Tempr
only) – NOT USED in latest f/w
R -100/+100 C K
61 Spare
62 Spare
63 Hours Of Operation R 0-65535 A
65 Set Clock HRS W 00-23 J
66 Set Clock MIN W 00-59 J
67 Set Clock DAY W 01-31 J
68 Set Clock MONTH W 01-12 J
69 Set Clock YEAR W 00-99 J
70 %O2 without HSR R -199.00-199.99% G
71 %O2 with HSR R -199.00-199.99% G
72 Cell Temperature R -99.9-99.9 or equiv in F or K F
73 PCB temperature (from MSP) R -99 to 99 C or equiv in F or K K
74 Atmos pressure R 0-1500mBar A
75 mA1 Input in % (comp signal) R 0.0-100.0% F
76 mA2 Input (ext sensor signal) R See Reg Details F
77 Status Flags register R 0-65535 I
78 Clock HOURS/MIN R 00-23 / 00-59 J
79 Clock SEC/DAY R 00-59 / 01-31 J
80 Clock MONTH/YEAR R 01-12 / 00-99 J
81 %O2 MINIMUM (stats) R -199.00-199.99% G
82 %O2 MAXIMUM (stats) R -199.00-199.99% G
83 VCOMP R 0-8191 A