Technical Data
thermoIMAGER TIM
Vibration, sinus shaped – testing Fc (acc. IEC60068-2-6)
Frequency range 10 - 500 Hz
Acceleration 29.42 m/s
2
(3 g)
Frequency change 1 octave/ min
Number of axes 3
Duration 1:30 h (3 x 0.30 h)
Vibration, broadband noise – testing Fh (acc. IEC60068-2-64)
Frequency change 10 - 2000 Hz
Acceleration, effective 39.3 m/s
2
(4.01 g
RMS
)
Frequency spectrum 10 - 106 Hz 0.9610 (m/s
2
)
2
/Hz (0.010 g
2
/Hz)
106 - 150 Hz +6 dB/ octave
150 - 500 Hz 1.9230 (m/s
2
)
2
/Hz (0.020 g
2
/Hz)
500 - 2000 Hz -6 dB/ octave
2000 Hz 0.1245 (m/s
2
)
2
/Hz (0.00126 g
2
/Hz)
Number of axes 3
Duration 3 h (3 x 1 h)