Technical Data
thermoIMAGER TIM
Model TIM M-1 M-05
Temperature range
(scalable)
450
1
… 1800 °C (27 Hz mode)
500
1
… 1800 °C
(80 Hz-Modus and 32 Hz mode)
600
1
…1800 °C (1 kHz mode)
900 … 2450 °C (27 Hz mode)
950 … 2450 °C (80 Hz and 32 Hz mode)
1100 … 2450 °C (1 kHz mode)
Spectral range 0.85 - 1.1 μm 500 - 540 nm
Detector CMOS,
764 x 480 Pixel@32 Hz
382 x 288 Pixel@80 Hz (switchable to 27 Hz)
72 x 56 Pixel@1 kHz (1 ms real-time analog output (0 - 10 V) from 8x8 pixel
(freely selectable)
764 x 8 Pixel@1 kHz (fast line scanning-mode, 1 ms real-time analog output
(0 - 10 V) from 8x8 Pixel (freely selectable)
Lenses (FOV) FOV@382x288 px:
20 ° x 15 °, 13 ° x 10 °, 7 ° x 5 °, 4 ° x 3 °
FOV@764x480 px:
39 ° x 25 °, 26 ° x 16 °, 13 ° x 8 °, 9 ° x 5 °
FOV@382x288 px:
13 ° x 10 °
FOV@764x480 px:
26 ° x 16 ° (F-number = 1.4)
System accuracy
2
For object temperature < 1400 °C:
±1 % of reading for 27/32/80 Hz
±1.5 % of reading for 1 kHz
For object temperature < 1600 °C:
±2 % of reading for 27/32/80 Hz
±2.5 % of reading for 1 kHz
For object temperature < 2000 °C:
±1 % of reading for 27/32/80 Hz
±1.5 % of reading for 1 kHz
For object temperature > 2000 °C:
±2 % of reading for 27/32/80 Hz
±2.5 % of reading for 1 kHz
Temperature
resolution (NETD)
3
< 2 K (< 900 °C)
< 4 K (< 1400 °C)
< 2 K (< 1400 °C)
< 4 K (< 2100 °C)
Warm-up time 10 min
Emissivity 0,100...1,100
Software TIMConnect
1) 2 +75 °C start temperature for optics with focal length f = 50 mm and f = 75 mm
2) At an ambient temperature of 25 °C
3) Specified NETD value applies to all frequencies.