Sample calibration certificate: A formatFigure A-1:
DENSITY CALIBRATION CERTIFICATE
CDM100PA18MBAZ00EANZZ 2012.10031Model Code:
Customer Tag:
Serial No:
Calibration Date:
Pressure Test:
2013-07-01 11:05:22
Calibration ID: 1.29825731
3263 psi(g)
A1 =
2.803826E-004
Time Period Temperature Compensation Coefficient:
Density Calibration Coefficients:
Case Temperature Compensation Coefficient:
Pressure Compensation Coefficient:
A2 = -2.549393E+007
A3 = 3.083073E-005
A4= 3.21
A5=
1.207270E-001
Calibrated Range = 0.300-1.300 g/cm³ (300-1300 kg/m³)
Flow Compensation Coefficient:
A6= 3.23
Time Period, TP secμ
Density (g/cm³)
Density Calibration Data:
5065.20540.00097
5587.31510.79882
5857.00311.24079
Performance Check:
Error (g/cm³)Density (g/cm³)
0.99822 -0.00007
Known Density Verification Data:
765.1Verification Time Period (Air) @ 20°C = μsec
CALIBRATED BY:
Density
Time period
Pressure
Definitions:
Temperature
g/cm³
µsec
psi(g)
°C
"A" Coefficients
Form Rev: 1
2013-08-29 13:55:05Printed Page 1 of 3
All equipment used for this calibration is calibrated at routine intervals against
standards that are traceable to the International System of Units (SI).
Emerson Process Management Micro Motion, Inc. 7070 Winchester Circle, Boulder, CO 80301
SAMPLE
Calibration certificate