Sample calibration certificate: K format, US unitsFigure A-3:
DENSITY CALIBRATION CERTIFICATE
CDM100PA18MBAZ00EANZZ 2012.10031Model Code:
Customer Tag:
Serial No:
Calibration Date:
Pressure Test:
2013-07-01 11:05:22
Calibration ID: 1.29825731
3263 psi(g)
7835|7845 "K" Coefficients (US Units)
K0 =
368
Density Calibration Coefficients @68°F
Temperature Compensation Coefficients:
Pressure Compensation Coefficients:
K18 = 371
K19 =
372
K20A=
373
Calibrated Range = 0.3-1.3 g/cm³ (300-1300 kg/m³)
K1 = 369
K2 = 370
K20B=
K21A=
K21B=
Range: 1-580 psi(g) Range: 435-1015 psi(g) Range: 870-1450 psi(g) Range: 1450-2175 psi(g)
373 373 373
0.0 0.0 0.0 0.0
0.0 0.0 0.0 0.0
0.0 0.0 0.0 0.0
Time Period, TP secμ
Density (g/cm³)
Density Calibration Data:
5065.20540.00097
5587.31510.79882
5857.00311.24079
Performance Check:
Error (g/cm³)Density (g/cm³)
0.99822 -0.00008
Known Density Verification Data:
765.1Verification Time Period (Air) @ 68°F = μsec
CALIBRATED BY:
D = Density (uncompensated)
TP = Time period
P= Pressure
Definitions:
°F
g/cm³
g/cm³Dt = Density (Temperature compensated)
Dp = Density (Pressure & Temperature compensated) g/cm³
µsec
T = Temperature
psi(g)
Form Rev: 1
2013-08-29 13:55:06Printed Page 3 of 3
All equipment used for this calibration is calibrated at routine intervals against
standards that are traceable to the International System of Units (SI).