Portaflow 220 User Manual 29
(Issue 1.0)
5: Troubleshooting
5.1 Overview
If you have a problem with your flow monitoring system it can be due to any of the following:
Faulty instrument
If you suspect the instrument is faulty you can check it out using a test block as described in Paragraph 5.4
.
This will establish that the instrument is functional and receiving a healthy signal from the connected
transducers.
Incorrect setup
A low, or zero, signal could be caused by incorrect set-up such as:
Incorrect site data entered into the instrument.
Incorrect or non-matching ultrasonic transducers selected for use.
Incorrectly fitted transducers – lack of couplant applied, incorrect spacing, insecure attachment.
Poor connections between the probes and the instrument.
Application problem
If you are certain that the instrument is healthy and suitably set-up for the current site; and the probes are
properly assembled and fitted correctly, there could be an application problem concerned with the site.
Check such conditions such as:
Poor pipe outer surface quality
Uneven surface preventing good surface contact with the transducer.
Flaking paint (should be removed).
Variable air gap in concrete-covered pipes affecting the ultrasonic signal quality.
Poor internal pipe construction
Rough internal pipe walls affecting fluid flow (see roughness factor).
Internal welds positioned in the transducer signal path affecting the signal quality.
The ‘drippings’ in galvanised-dipped pipes or other irregularities interfering with the signal path.
Incorrect probe location
Transducers located too close to bends or valves, disturbing the flow profile.
Transducers located too close to insertion probes, disturbing the flow profile.
For horizontal pipework transducers should not be positioned on the top of the pipe.
Poor fluid conditions within the pipe
Fluid contains bubbles, high particle density or sludge.
Air in the top of the pipe.
Low fluid flow within the pipe
Pipe obstructions.
Malfunctioning valve not opening fully (or closed inadvertently).
Liquid content problems
Multiple liquid contents do not comply accurately to expected sound speed criteria.
Very hot pipe almost turns water to steam and therefore exhibits the wrong speed characteristics –
could be due to reduced pipe pressure.
Flashover – liquid turns into a gas because of lower than required pressure.