EasyManuals Logo

Micronics portaflow 220 User Manual

Micronics portaflow 220
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #36 background imageLoading...
Page #36 background image
5: Troubleshooting
32 Portaflow 220 User Manual
(Issue 1.0)
4-20mA ERRORS
mA out > Max Interpretation: The actual flow is higher than the maximum set on the mA range.
Response: Re-scale the 4-20mA output to be able to cope with the higher flow –
see Paragraph 3.7.2
.
Calibration 20mA
Error!
NOTE: The 4-20mA output is calibrated before the instrument leaves the factory and
should not require further adjustment.
Interpretation: You have adjusted the DAC outside its accepted range when
calibrating the 20mA signal output.
Response: Re-calibrate the 4-20mA output – see Paragraph 3.7.2
.
Calibration 4mA Error! NOTE: The 4-20mA output is calibrated before the instrument leaves the factory and
should not require further adjustment.
Interpretation: You have adjusted the DAC outside its accepted range when
calibrating the 4mA signal output.
Response: Re-calibrate the 4-20mA output – see Paragraph 3.7.2
.
SET-UP ERRORS
Pipe OD out of range Interpretation: You have entered an out-of-range value for the pipe outside
diameter dimension – i.e. larger or smaller than the unit or sensor can be used on.
Response: Enter a valid number.
Wall thickness out of
range
Interpretation: You have entered an out-of-range value for the pipe wall thickness
dimension – accepted range is 1mm - 75mm.
Response: Enter a valid number.
Lining thickness out of
range
Interpretation: You have entered an out-of-range value for the lining thickness
dimension – acceptable range is 0mm - 10mm.
Response: Enter a valid number.
Temperature range Interpretation: You have entered an out-of-range value for the fluid Temperature.
Accepted temperature range -20°C to +300°C.
Response: Enter a valid number.
Invalid Date or Time Interpretation: The entered Date or Time is invalid, or when setting up ‘timed’ data
logging the Stop time is set earlier than the Start time.
Response: Enter a valid Date and Time.
Sensors: INVALID Interpretation: The selected temperature is higher than the maximum allowed for
the sensor type.
Response: Enter a different temperature.
Mode: Err Typ Interpretation: The selected sensors are invalid and the mode cannot be verified.
Response: Select a mode that gives a non-zero separation distance.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Micronics portaflow 220 and is the answer not in the manual?

Micronics portaflow 220 Specifications

General IconGeneral
BrandMicronics
Modelportaflow 220
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals