DOC 6501_Release V
ATS-6501 Users Guide
26
shall have the ability to pad a sequence of data bits with zero bits by specifying a bit
count larger than the number of data bits in the file.
The following example loads binary file with an 8,000,000 bit count to slot 5, block 1.
ATS-6501>bitcode_memblock --file 1m_othdrtrl.bin --format binary --bitcount 8000000 5 1
[OK] 2014-04-29T16:17:53Z
To verify the settings use the settings hardware:slots:5:tsc4391:memblocks command (where
N=Slot #) to obtain the current information on the memory blocks for the 4391 card.
ATS-6501>settings hardware:slots:5:tsc4391:memblocks
[1]
[bit_count] 8000000
[file_format] binary
[file_name] 1m_othdrtrl.bin
[2]
[bit_count] 0
[file_format] binary
[file_name]
[3]
[bit_count] 0
[file_format] binary
[file_name]
[4]
[bit_count] 0
[file_format] binary
[file_name]
[OK] 2014-04-29T23:18:13Z
Streams: This command shall configure one or more bit generator streams on a 4391A output
card.
To set the Stream parameters use the bitcode_stream command and specify the clock (1 | 2), the
data (block (memory block #) | square), the slot # (1-6), and a comma delimited stream list. Users
can enable/disable the stream using the --enable 0 (off) or 1 (on).
Syntax
ATS-6501> bitcode_stream --clock CLOCK --data <BLOCK|square> --enable <true|false> SLOT
[STREAMLIST]
o clock: Bit generator clock number to assign to the specified stream(s). CLOCK must be
1 or 2. If not specified the assigned clock number(s) shall remain unchanged.
o data: Data source to assign to the specified stream(s). If SOURCE is an integer between
1 and 4, the specified memory block shall be assigned as the data source. If SOURCE is
the text string square”, the FPGA’s square wave generator shall be assigned as the data
source. If not specified, the assigned data source shall remain unchanged. If the
specified data source is not an integer between 1 and 4 or the text string “square” the
following error message shall be displayed: “Data must be ‘square’ or a block number
between 1 and 4”.