MI104 Weighing Indicator User Manual
You can calculate new rate by this formula & enter it:
New rate = Old rate*Display weight/ Calibration weight.
This display will show XXXXX for indicating the internal
counts.
Normal weighing mode. (check weighing, accumulation)
Animal weighing mode. (scale can lock reading, when little
unstable)
This is a subtraction scale (print out “-“ weight)
Peak Hold mode. (Scale can hold maximum reading)
7. CALIBRATION
In weighing mode, press key and key together.
Press repeatedly until display shows.
Press , display will show.
Enter the password. Press , and
Display will show
Press , display will show.
Enter the function by pressing twice , display will show
Press repeatedly until display shows.